Company record
Maquet Cardiopulmonary Gmbh
0 FDA warning letters since 2023 and 4 recalls in the last 12 months.
Get an email the day FDA posts anything new about Maquet Cardiopulmonary Gmbh.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Sep 23 | Class II | Maquet Cardiopulmonary Gmbh Device |
Percutaneous Insertion Kit (PIK), Guidewire 100 cm, Model Numbers: 701053068 (USA), 701047384 (OUS); The packaging may not consistently maintain sterile barrier integrity. |
| Sep 23 | Class II | Maquet Cardiopulmonary Gmbh Device |
Percutaneous Insertion Kit (PIK), Guidewire 150 cm, Model Numbers: 701053069 (USA), 701047385 (OUS); The packaging may not consistently maintain sterile barrier integrity. |
| Sep 23 | Class II | Maquet Cardiopulmonary Gmbh Device |
Percutaneous Insertion Kit (PIK), dilator set L, Model Numbers: 701054427 (USA), 701054427 (OUS); The packaging may not consistently maintain sterile barrier integrity. |
| Feb 11 | Class II | Maquet Cardiopulmonary Gmbh Device |
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720; Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised. |